Where the measuring surface is coated and it is required to eliminate the coating thickness from the results, a suitable instrument that uses mode 3 should be used.

Where it is required to find the thinnest point within a given area, a scanning should be performed. For this purpose an А-scan instrument should be used.

Where many readings are to be recorded, an instrument with data logging facility should be considered.

A.2.2 Probes

Probe selection depends on equipment type, material thickness, surface condition and coating condition.

For digital display instruments the probes as specified by the manufacturer should be used. For А-scan instruments the following guidelines can be applied:

  • the probe frequency should be selected such that it produces at least 1,5 times the wavelength in the test object (see 8.2.2);

  • generally, single transducer probes should be used for thicknesses of 10 mm and above. The multiple echo technique (mode 3) should only be used with single transducer probes;

  • where thickness is below 10 mm, double transducer probes may be applied;

  • if the thickness is expected to be below 5 mm, dual element probes with special focal range should be used;

  • when the object is curved, consideration should be given to the selection of probe diameter;

  • on a coated object a single transducer probe should be used with mode 3 to allow compensation for the coating thickness.

A.2.3 Setting of the instrument

Setting of the instrument is done on a step-wedge with a thickness range covering the expected range of the object.

Material and temperature shall be equivalent to the object.

A.2.4 Measuring

Where several back-wall echoes may be read (only single transducer technique), the most accurate results are achieved by reading the nth echo and dividing the reading with n. Where this technique is used on a coated surface, the distance from echo no. 1 to echo no. n is read and divided by и-1. Hereby the coating thickness is not included in the result.

Where only one back-wall echo is used, the reading should be taken in the same position of the echo as the reading during setting of the instrument. If the surface is coated, the coating thickness multiplied with the sound velocity ratio metal/coating is included in the reading and should be subtracted before recording the result.

Where high reproducibility is essential, the exact position of the measuring point is to be documented or assured in another way. Where it is essential to detect the thinnest point within a given area, scanning should be performed. This will normally require an А-scan instrument (type 5.1 b) or type 5.1 c)).

Use of digital display instruments should strictly follow the manufacturer's instructions.

Unexpected measurements may be due to internal discontinuities. These should be verified by supplementary investigations, e.g. by using angle-beam probes.

A.3 Measurement of corrosion with pitting

A.3.1 Instrument

For thickness measurement where pitting may be expected, an А-scan instrument should be used (type 5.1 b) or type 5.1 c)).

A.3.2 Probes

For detection of pitting, a dual element probe is the most suitable. The selected probe should have a focal distance corresponding to the expected distance to the pitting.

A.3.3 Setting of the instrument

The setting of the instrument is done on a step-wedge with a thickness range covering the expected range of the object. Material and temperature shall be equivalent to the object. Where small diameter pitting is expected, detection sensitivity is to be verified on a calibration block with small diameter flat-bottom holes in the same distance range as the expected pitting.

A.3.4 Measuring

When searching for pitting, only the first back-wall echo should be used. Echoes from pitting may occur together with the back-wall echo.

Where reflector type cannot be identified as either corrosion or inclusion, supplementary investigation should be carried out using angle probes. 45° angle probes are specially suited for differentiating between inclusions and pitting.

Table A.1 — Corrosion in steel - Recommended ultrasonic techniques

No.

Description

Typical corrosion origin and mechanism

Illustration

Recommended ultrasonic technique

1

Uniform corrosion

Occurs in corrosive environments such as:

  • water saturated with oxygen;

  • sour solutions;

  • condensed water from wet gas.


A.2

Development in uniform corrosion

2

Pitting

Corroded areas have clear limits while the adjacent areas are typically not attacked.

Pitting can take different shapes, depending on structure and texture of the material, and on surface condition.






A.3







Type A: Flat pitting


Type C: Semicircular pits




Ґ Л '



Type B: Undermining pitting


Type D: Surface breaking spherical



Table A.1 — Corrosion in steel - Recommended ultrasonic techniques (continued)

No.

Description

Typical corrosion origin and mechanism

Illustration

Recommended ultrasonic technique

2a

Pitting

Distribution patterns








See Note



• •

• •


• •

• • e









3

Deposit corrosion, Crevice corrosion

Occurs under deposits and in narrow water-filled crevices





1


See Note












Table A.1 — Corrosion in steel - Recommended ultrasonic techniques (continued)

No.

Description

Typical corrosion origin and mechanism

illustration

Recommended ultrasonic technique

4

Galvanic corrosion

Dissimilar metals






See Note










—(




5

Flow-induced corrosion


."Il

_ і

і

4||l


See Note



Table A.1— Corrosion in steel - Recommended ultrasonic techniques (continued)

No.

Description

Typical corrosion origin and mechanism

Illustration

Recommended ultrasonic technique

6

Turbulence corrosion


4 'i

IL

1 1

See Note

7

Mesa-type corrosion



See Note

8

Cavitation corrosion


1 ЙЖ

See Note



  1. 1 — Corrosion in steel - Recommended ultrasonic techniques (concluded)

No.

Description

Typical corrosion origin and mechanism

Illustration

Recommended ultrasonic technique

9

Weld zone corrosion



See Note

NOTE These corrosion forms are shown to illustrate the possibilities and difficulties that can be encountered when achieving to detect and quantify corrosion. The illustrations are for information only. A specific recommendation regarding the technique to be applied for each case cannot be given, as it would depend on the access conditions, material thicknesses and other parameters.



Annex В

(informative)

Instrument settings

  1. 1 — Instrument setting on a reference block with multiple steps

SELECT REFERENCE

BLOCK ...

OF SAME MATERIAL AND SAME SURFACE CONDITION

OF SAME MATERIAL AND OTHER SURFACE CONDITION

OF DIFFERENT MATERIAL AND SAME SURFACE CONDITION

OF DIFFERENT MATERIAL AND OTHER SURFACE CONDITION

CALIBRATE THE EQUIPMENT

Calibrate on a thickness above and below the thickness range to be measured.

Calibrate on a thickness above and below the thickness range to be measured.

Calibrate on a thickness above and below the thickness range to be measured.

Calibrate on a thickness above and below the thickness range to be measured.

VERIFY LINEARITY AT INTERMEDIATE STEPS ...

if more than 2 steps available

if more than 2 steps available

if more than 2 steps available

If more than 2 steps available

CORRECT THE SETTING

Not necessary

Check and correct zero setting on the test object

Re-calibrate on the test object if possible

or use known velocity to correct the reading.

Re-calibrate on the test object if possible or check and correct zero setting on the test object and use known velocity value.

UNCERTAINTY OF MEASUREMENT RELATED TO SETTING OF THE INSTRUMENT DEPENDS ON ...

Accuracy of reference block thicknesses and, if only 2 steps used, uncertainty of linearity

Accuracy of reference block thicknesses and the surface condition of the test object and, if only 2 steps used, uncertainty of linearity

Accuracy of reference block thicknesses and

accuracy of thicknesses of the test object or

validity of known value of velocity

and, if only 2 steps used, uncertainty of linearity

Accuracy of reference block thicknesses and accuracy of thicknesses of the test object and

the surface condition of the test object or

validity of known value of velocity

and, if only 2 steps used uncertainty of linearity



  1. 2 — Instrument setting on a reference block with one thickness or without a reference block

REFERENCE BLOCK

OF SAME MATERIAL AND SAME SURFACE CONDITION

OF SAME MATERIAL AND OTHER SURFACE CONDITION

NO REFERENCE BLOCK OF SAME MATERIAL AVAILABLE

CALIBRATE THE EQUIPMENT

Set the velocity and zero to agree with the known value and thickness

Set the velocity and zero to agree with the known value and thickness

Set the velocity to a known value for the test object

and

Set zero by using a known value or by using mode 3 or by using automatic probe recognition

VERIFY LINEARITY AT INTERMEDIATE STEPS

Not possible

Not possible

Not possible

CORRECT THE SETTING

Not necessary

Check and correct zero setting on test object

Not possible

UNCERTAINTY OF MEASUREMENT RELATED TO SETTING OF INSTRUMENT DEPENDS ON

Accuracy of reference block thickness and uncertainty of linearity

Accuracy of reference block thickness and uncertainty of linearity and surface condition of the test object

The validity of the known values

Annex C

(informative)

Parameters influencing accuracy

C.1 Parameters influencing accuracy

The following Table C.1 lists the parameters influencing accuracy:

  1. 1 — Table of parameters influencing accuracy

Item

Parameter

Result

Possible improvements

Test object

Material

Composition

Attenuation, absorption, scattering and local variation of velocity

Setting of instrument on the same material as test object

Structure

Anisotropy

Surface condition

Cleanliness

Local variations of surface conditions lead to variations of couplant thickness

Cleaning

Roughness

Grind surface as required

Surface profile

Using of small diameter probe

Coating

Coating

Coating velocity different from base material velocity resulting in inaccuracy

Removing coating or using mode 3

Paint

Surface treatment

Geometry

Non-parallelism

Back-wall echo can disappear or can be distorted

Parallelism should be within the probes beam divergence angle (± 1,22 arcsin A/d)

Curvature

Loss of coupling efficiency

Use a smaller diameter probe

Range

Distortion of back-wall echo caused by attenuation

Using mode 1 and a lower probe frequency using mode 4



(continued)



Table C.1 — Table of parameters influencing accuracy (concluded)

Reference

Method

Uncertainty of calibration method

Inaccurate readings

Using block representative of part, steps thinner and thicker than expected thickness, choice of calibration method see Annex В

Reference block

Thickness and velocity uncertainty

Accuracy cannot be better than block uncertainties

Accurate measurement of block thickness and sound velocity

Measuring

Equipment

Resolution

Accuracy cannot be better than system resolution

Using higher accuracy instrument, higher probe frequency and broadband probes

Cable length

Excessive cable length distorts the signals

Using shorter cable and calibrate with the same cable

Drift of instrument

Inaccurate readings

Warming-up the unit and wait for stable reading or use stable equipment

Time of flight

Accuracy cannot be better than time of flight measurement accuracy

Using higher accuracy instrument

Linearity

Inaccurate readings

Ensuring linearity of system

Trigger point

Inaccurate readings

Selecting best trigger point

Operation

V-path

Wrong reading because thickness differs from ultrasonic path

Using a thickness gage with V-path correction or taking into account the roof angle and separation

Using a single element probe

Phase shift

Erroneous reading

Taking the phase shift into account

Repeatability

Use of unit

Method

Improper operation

Providing correct procedure or instructions. Conducting repeatability tests

Coupling

Bad coupling introduces dispersion in the readings.

Selecting couplant to suit the surface conditions

Using mode 3 if possible

User training

Error on reading

Operator training

Miscellaneous

Temperature

Variation of sound

I velocity

Error on reading

Calibrating at the same temperature as test object or correcting calibration for change of sound velocity